    ALERT

    If unable to palpate an artery because of a weak pulse, use an ultrasonic stethoscope (Figure 1)Figure 1.

    Blood pressure (BP) measurements with an electronic BP device may be affected by excessive movement, such as with seizures, tremors, or shivering, and irregular heart rates.

    OVERVIEW

    BP is the force exerted by blood against the vessel walls. During a normal cardiac cycle, BP reaches a peak, followed by a trough. The peak pressure occurs when the heart’s ventricular contraction, or systole, forces blood under high pressure into the aorta. When the ventricles relax, the blood remaining in the arteries exerts a trough, or diastolic, pressure against the arterial wall. Diastolic pressure is the minimum pressure exerted against the arterial wall.

    Patients at risk for alterations in BP measurement include those who have:

    • Circulatory shock (hypovolemic, septic, cardiogenic, or neurogenic)
    • Acute or chronic pain
    • Rapid IV infusion of fluids or blood products
    • Increased intracranial pressure
    • Postoperative status
    • Preeclampsia of pregnancy

    The standard unit for measuring BP is millimeters of mercury (mm Hg). The measurement indicates the height to which BP can sustain the column of mercury.

    The most common methods for measuring BP are auscultation, using a sphygmomanometer and a stethoscope, and measurement using an electronic BP monitor. Palpation may be used to obtain an estimate of systolic BP before using the auscultation method.

    During auscultation, as the sphygmomanometer cuff is deflated, five different sounds, called Korotkoff sounds, are heard over the artery. Each sound has a distinct characteristic (Figure 2)Figure 2. BP is recorded with the systolic reading (first Korotkoff sound) before the diastolic reading (beginning of the fifth Korotkoff sound). The difference between systolic pressure and diastolic pressure is the pulse pressure. For a BP of 120/80 mm Hg, the pulse pressure is 40 mm Hg, the difference between 120 mm Hg and 80 mm Hg.

    Cuff size should be proportionate to the limb circumference (Table 1)Table 1. Most adults require a large adult cuff. The correct cuff size is especially important for obtaining accurate readings in pediatric patients and patients with obesity.undefined#ref7">7 An improperly sized cuff produces an inaccurate BP measurement. Using a cuff that is too narrow results in an overestimation of BP, whereas using a cuff that is too wide results in an underestimation of BP.1,6

    When measuring BP in the upper arm is not possible—for example, when the available BP cuffs do not fit the upper arm properly—BP may be measured in the forearm. To obtain the most accurate reading, the health care team member must use the proper-size BP cuff for the forearm, which typically has a smaller circumference than the upper arm. BP measurements in the forearm and upper arm are not interchangeable. Systolic BP readings tend to be higher in more distal arteries.9

    SUPPLIES

    See Supplies tab at the top of the page.

    EDUCATION

    • Provide developmentally and culturally appropriate education based on the desire for knowledge, readiness to learn, and overall neurologic and psychosocial state.
    • Explain the equipment and the procedure to the patient and family.
    • Instruct the patient and family about ambulatory BP threshold guidelines. In adults, normal BP is less than 120/80 mm Hg (Table 2)Table 2.3,4
    • Educate the patient and family about the risk factors for hypertension.
      • Family history of hypertension, premature heart disease, lipidemia, or renal disease
      • Obesity
      • Cigarette smoking
      • Heavy alcohol consumption
      • High blood cholesterol and triglyceride levels
      • Prolonged stress from psychosocial and environmental factors
      • Sedentary lifestyle
    • Educate the patient and family regarding the primary strategies for preventing hypertension.2
      • Managing weight
      • Engaging in daily exercise
      • Limiting sodium and saturated fat in the diet
      • Taking medications as prescribed
      • Limiting alcohol intake
      • Maintaining adequate intake of dietary potassium and calcium
      • Avoiding tobacco products
    • Instruct the patient and family to ensure that the patient has adequate rest before BP measurements and that BP measurements are performed at the same time each day using the same limb with the patient in the same position, either sitting or lying down.
    • Explain that the patient must remain still and quiet during the procedure.
    • Explain that Korotkoff sounds may be difficult to hear for one of the following reasons (Table 3)Table 3.
      • Cuff is too loose, not big enough, or too narrow.
      • Stethoscope is not over the arterial pulse.
      • Cuff is deflated too quickly or too slowly.
      • Cuff is not pumped high enough for systolic readings.
    • Encourage questions and answer them as they arise.

    ASSESSMENT AND PREPARATION

    Assessment

    1. Perform hand hygiene before patient contact. Don appropriate personal protective equipment (PPE) based on the patient’s need for isolation precautions or the risk of exposure to bodily fluids.
    2. Introduce yourself to the patient.
    3. Verify the correct patient using two identifiers.
    4. Review the patient’s medical record for a history of hypertension, cardiovascular disease, renal disease, diabetes, and other factors that influence BP (e.g., weight, smoking, medications).
    5. Assess the patient for risk factors for BP alterations.
      1. Circulatory shock (hypovolemic, septic, cardiogenic, or neurogenic)
      2. Acute or chronic pain
      3. Rapid IV infusion of fluids or blood products
      4. Increased intracranial pressure
      5. Postoperative status
      6. Preeclampsia of pregnancy
    6. Assess the patient for signs and symptoms of BP alterations.
      1. Assess a patient at risk for high BP for the following signs and symptoms.
        1. Headache (usually occipital)
        2. Facial flushing
        3. Nosebleed
        4. Fatigue
      2. Assess a patient at risk for low BP for the following signs and symptoms.
        1. Dizziness
        2. Mental confusion
        3. Restlessness
        4. Pale, dusky, or cyanotic skin and mucous membranes
        5. Cool, mottled skin over the extremities
    7. Determine the best site for BP assessment. Avoid applying the cuff to an extremity in the following situations.
      1. IV fluids are infusing.
      2. An arteriovenous shunt or fistula is present.
      3. An ipsilateral breast or axillary surgery has been performed. The postsurgical side may be used if other sites are not available or practical, as the risk of developing lymphedema from occasional BP measurement following breast or lymph surgery is low.5
      4. The extremity has been traumatized.
      5. The extremity has known infections or medical conditions (e.g., those causing vasoconstriction or a tumor pressing on the vascular supply).
      6. The extremity has a cast or bulky bandage.
    8. Determine the previous baseline BP and measurement site, if available, from the patient’s record.
    9. Determine the appropriateness of using an electronic BP measurement (Box 1)Box 1.

    Preparation

    1. Make sure the patient has not exercised, ingested caffeine, or smoked for 30 minutes before BP assessment.6 Make sure the patient does not have to void.
      Rationale: A full bladder increases pressure on the kidneys and can increase blood pressure.
    2. Make sure the room is warm (approximately 22°C to 23°C [71.6°F to 73.4°F]).6
      Rationale: Exposure to cold can increase systolic BP.6
    3. Measure upper extremity circumference and select the appropriate-size cuff (Figure 3)Figure 3 (Table 1)Table 1.
      Rationale: An improperly sized cuff produces inaccurate BP measurements.
    4. Tell the patient that BP is going to be taken and that the cuff will squeeze the arm.

    PROCEDURE

    Auscultation Method

    1. Perform hand hygiene and don gloves. Don additional PPE based on the patient’s need for isolation precautions or the risk of exposure to bodily fluids.
    2. Verify the correct patient using two identifiers.
    3. Explain the procedure and ensure that the patient agrees to treatment.
    4. Have the patient sit or lie down. Record the patient’s position when performing orthostatic vital signs. If the patient is sitting, ensure the patient’s back is supported.
      Rationale: BP is generally higher in the supine position than in the sitting position.6
    5. Position the patient’s forearm at heart level, supported, with the palm facing up (Figure 4)Figure 4. If the patient is sitting, instruct the patient to keep the feet flat on the floor with the legs uncrossed. If the patient is supine, ensure the legs are not crossed.
      Rationale: If the patient’s arm is not supported at the heart level, a lower BP will be recorded when the arm is above heart level, and a higher BP will be recorded when the arm is below heart level.6 Leg crossing can artificially increase systolic and diastolic BP.
    6. Expose the patient’s arm fully by removing any constricting clothing. Do not place the BP cuff over clothing.
      Rationale: Placing the cuff over clothing may affect the BP measurement.
    7. Apply the BP cuff. Do not place the cuff over a bony prominence, nerve, or joint.6
      1. Upper arm
        1. Palpate the brachial artery for a pulse (Figure 5A)Figure 5A.
        2. Position the cuff above the antecubital fossa (Figure 5B)Figure 5B.
        3. Apply the cuff in the middle of the patient’s arm approximately midpoint with the patient’s sternum, and center the arrows marked on the cuff over the brachial artery (Figure 5C)Figure 5C.6 If the cuff has no center arrows, estimate the center of the bladder and place it over the artery.
          Rationale: Positioning the cuff bladder directly over the brachial artery ensures that proper pressure is applied during inflation.
        4. Ensure the fully deflated cuff is wrapped evenly and snugly around the patient’s upper arm.
          Rationale: A loose-fitting cuff can cause an artificially high reading.
      2. Forearm
        1. Palpate the radial artery for a pulse.
        2. Position the cuff below the antecubital fossa.
        3. Apply the cuff by centering the arrows marked on the cuff over the radial artery and below the antecubital fossa.6 If the cuff has no center arrows, estimate the center of the bladder and place it over the artery.
          Rationale: Positioning the cuff bladder directly over the radial artery ensures that proper pressure is applied during inflation.
        4. Ensure the fully deflated cuff is wrapped evenly and snugly around the patient’s forearm.
          Rationale: A loose-fitting cuff can cause an artificially high reading.
    8. Position the manometer vertically at eye level.
      Rationale: Looking up or down at the scale can result in an incorrect measurement.
    9. Ask the patient not to speak while BP is being measured.
    10. Optional step: Estimate the patient’s systolic BP by palpation.
      1. Locate the brachial or radial pulse.
      2. Palpate the artery distal to the cuff with the fingertips of the nondominant hand while inflating the cuff rapidly to a pressure above the point at which the pulse disappears.
      3. Slowly deflate the cuff and observe the point when the pulse reappears.
      4. Deflate the cuff fully.
    11. Place the stethoscope earpieces in the ears and make sure sounds are clear, not muffled.
    12. Locate the brachial or radial artery as appropriate and place either the bell or the diaphragm of the stethoscope6 over the artery below the bottom of the cuff.
      1. Do not let the diaphragm touch the cuff or clothing.
      2. Do not place the diaphragm under the cuff.
      3. Do not place excessive pressure on the stethoscope head.
      4. If unable to palpate the artery because of weakened pulse, use an ultrasonic stethoscope (Figure 1)Figure 1.
        Rationale: Proper stethoscope placement ensures the best sound reception. An improperly positioned stethoscope can cause muffled sounds that often result in artificially low systolic and artificially high diastolic readings.
    13. Turn the valve of the pressure bulb clockwise until tight. Quickly inflate the cuff above the patient’s estimated palpated systolic pressure, the patient’s previously documented systolic pressure, or the point at which Korotkoff sounds cease.
      Rationale: Closing the valve prevents air leak during inflation. Rapid cuff inflation ensures accurate measurement of systolic pressure.
    14. Slowly release the pressure bulb valve, allowing the manometer needle to fall slowly and continuously at a rate of 2 to 3 mm Hg per second.6
      Rationale: Too rapid or slow a decline in the mercury level can cause inaccurate measurements.
    15. Observe the point on the manometer at which the first Korotkoff sound is heard, indicating the systolic BP. The sound will slowly increase in intensity.
      Rationale: This first Korotkoff sound is a snapping sound. This sound, for at least two consecutive heartbeats, reflects the systolic BP.
    16. Continue to deflate the cuff gradually; observe the point on the manometer at which all Korotkoff sounds disappear, indicating the diastolic BP.
      Rationale: The fifth Korotkoff sound falls silent as the cuff pressure drops below the diastolic pressure. Thus, the beginning of the fifth Korotkoff sound indicates diastolic pressure in adults.6
    17. When the sounds disappear, quickly deflate the cuff completely.
    18. Remove the cuff from the patient’s arm unless a repeat measurement is needed.
      Rationale: Continuous cuff inflation can cause arterial occlusion, resulting in numbness and tingling of the arm.
    19. If this is the patient’s first BP assessment, repeat the procedure on the other arm. If there is a consistent difference between the BP in the patient’s arms, use the arm with the higher pressure per the organization’s practice or the patient’s condition.1,6
      Rationale: Comparison of BP in both arms helps detect cardiovascular, neurologic, and musculoskeletal abnormalities. A difference of greater than 10 mm Hg may be clinically significant.6
    20. Help the patient resume a comfortable position and return any removed clothing. Inform the patient of the BP reading, as appropriate.
    21. Report abnormal values to the practitioner (Table 2)Table 2.
    22. Clean the BP cuff per the manufacturer’s instructions and the organization’s practice. Clean the earpieces and diaphragm of the stethoscope with an alcohol swab or per the organization’s practice. Return the equipment to its assigned storage space.
    23. Discard supplies, remove PPE, and perform hand hygiene.
    24. Document the procedure in the patient’s record.

    Electronic Method

    1. Perform hand hygiene and don gloves. Don additional PPE based on the patient’s need for isolation precautions or the risk of exposure to bodily fluids.
    2. Verify the correct patient using two identifiers.
    3. Explain the procedure and ensure that the patient agrees to treatment.
    4. Have the patient sit or lie down. Record the patient’s position when performing orthostatic vital signs. If the patient is sitting, ensure the patient’s back is supported.6
      Rationale: BP is generally higher in the supine position than in the sitting position.6
    5. Plug in the device and place it near the patient’s extremity, then choose an appropriate-size cuff (Table 1)Table 1. Ensure that the BP cuff is appropriate for the electronic device per the manufacturer’s instructions.
    6. Prepare the BP cuff by manually squeezing all the air out of it and connecting it to the connector hose.
    7. Position the patient’s forearm, supported, at heart level with the palm facing up (Figure 4)Figure 4. If the patient is sitting, instruct the patient to keep the feet flat on the floor with the legs uncrossed. If the patient is supine, ensure the legs are not crossed.
      Rationale: If the patient’s arm is not supported at the heart level, a lower BP will be recorded when the arm is above heart level, and a higher BP will be recorded when the arm is below heart level.6 Leg crossing can artificially increase systolic and diastolic BP.
    8. Expose the upper arm fully by removing any constricting clothing. Do not place the BP cuff over clothing.
      Rationale: Placing the cuff over clothing may affect the BP measurement.
    9. Apply the BP cuff. Do not place the cuff over a bony prominence, superficial nerve, or joint.
      1. Upper arm
        1. Palpate the brachial artery for a pulse (Figure 5A)Figure 5A.
        2. Position the cuff above the antecubital fossa (Figure 5B)Figure 5B.
        3. Apply the cuff in the middle of the patient’s arm approximately midpoint with the patient’s sternum, and center the arrows marked on the cuff over the brachial artery (Figure 5C)Figure 5C.6 If the cuff has no center arrows, estimate the center of the bladder and place it over the artery.
          Rationale: Positioning the cuff bladder directly over the brachial artery ensures that proper pressure is applied during inflation.
        4. Ensure the fully deflated cuff is wrapped evenly and snugly around the patient’s upper arm.
          Rationale: A loose-fitting cuff can cause an artificially high reading.
      2. Forearm
        1. Palpate the radial artery for a pulse.
        2. Position the cuff below the antecubital fossa.
        3. Apply the cuff by centering the arrows marked on the cuff over the radial artery and below the antecubital fossa.6 If the cuff has no center arrows, estimate the center of the bladder and place it over the artery.
          Rationale: Positioning the cuff bladder directly over the radial artery ensures that proper pressure is applied during inflation.
        4. Ensure the fully deflated cuff is wrapped evenly and snugly around the patient’s forearm.
          Rationale: A loose-fitting cuff can cause an artificially high reading.
    10. Ask the patient not to speak while BP is being measured.
    11. Verify that the connector hose between the cuff and the machine is not kinked.
    12. Set the frequency control for automatic or manual and then press the start button following the manufacturer’s instructions.
    13. When the deflation is complete, observe the digital display, which provides the most recent values.
    14. Obtain additional readings by pressing the start button rather than the cancel button.
      Rationale: Pressing the cancel button immediately deflates the cuff.
    15. If frequent BP measurements are required, leave the cuff in place. Set the frequency of BP measurements and the upper and lower alarm limits for systolic, diastolic, and mean BP readings.
      1. Remove the cuff at regular intervals to assess the underlying skin integrity.
      2. If possible, alternate BP sites.
        Patients with abnormal bleeding tendencies are at risk for microvascular rupture from repeated inflations.
    16. If repeat measurements are not needed, remove the cuff from the patient’s arm.
      Rationale: Continuous cuff inflation can cause arterial occlusion, resulting in numbness and tingling of the arm.
    17. If this is the patient’s first BP assessment, repeat the procedure on the other arm. If there is a consistent difference between the BP in the patient’s arms, use the arm with the higher pressure per the organization’s practice or the patient’s condition.1,6
      Rationale: Comparison of BP in both arms helps detect cardiovascular, neurologic, and musculoskeletal abnormalities. A difference of greater than 10 mm Hg may be clinically significant.6
    18. Help the patient resume a comfortable position and return any removed clothing. Inform the patient of the BP reading, as appropriate.
    19. Report abnormal values to the practitioner (Table 2)Table 2.
    20. Clean the BP cuff and electronic device per the manufacturer’s instructions and the organization’s practice. Return the equipment to its assigned storage space.
    21. Discard supplies, remove PPE, and perform hand hygiene.
    22. Document the procedure in the patient’s record.

    MONITORING AND CARE

    1. If assessing BP for the first time, establish the BP reading as the baseline if it is within the acceptable range.
    2. Compare the BP reading with the patient’s previous baseline and the usual BP for the patient’s age.
    3. Assess, treat, and reassess pain.

    EXPECTED OUTCOMES

    • BP is within acceptable range for patient’s age and body size.
    • Patient tolerates procedure.

    UNEXPECTED OUTCOMES

    • BP is above acceptable range.
    • BP is below acceptable range or insufficient for adequate perfusion and oxygenation of tissues.
    • BP reading cannot be obtained.
    • Patient experiences orthostatic hypotension.
    • A significant difference exists between left arm and right arm BP readings.

    DOCUMENTATION

    • BP measurement
    • Method
    • Site assessed and patient’s position
    • BP measurement after administration of specific therapies
    • Signs and symptoms of BP alterations
    • Abnormal findings
    • Unexpected outcomes and related interventions
    • Education

    PEDIATRIC CONSIDERATIONS

    • The right arm is preferred for BP measurement in pediatric patients for consistency and comparison with standardized BP measurement tables for age and weight.8
    • BP measurement may frighten pediatric patients. Prepare the patient for the squeezing feeling of the inflating BP cuff by comparing the sensation to an elastic band on the finger or a tight hug on the arm.
    • Measure the patient’s BP before performing anxiety-producing tests or procedures.
    • Assess the BP of a pediatric patient with respect to body size and age.
      • Heavier and taller patients have a higher BP than smaller patients of the same age.
      • During adolescence, BP continues to vary according to body size.
      • The normal range for 10- to 17-year-old patients at the 90th percentile for weight is systolic 124 to 136 mm Hg and diastolic 77 to 84 mm Hg for biological male patients and systolic 124 to 127 mm Hg and diastolic 63 to 74 mm Hg for biological female patients.6
      • Korotkoff sounds are difficult to hear in pediatric patients because of their low frequency and amplitude. A pediatric stethoscope bell is often helpful.
    • Although the beginning of the fifth Korotkoff sound indicates diastolic pressure in adults, the fourth Korotkoff (distinct muffling) indicates diastolic pressure in pediatric patients.

    OLDER ADULT CONSIDERATIONS

    • Older adults, especially frail older adults, typically have lost upper arm mass, requiring special attention to BP cuff size.
    • The skin of older adults is more fragile and susceptible to damage from cuff pressure when BP measurements are frequent. More frequent assessment of the skin under the cuff or rotation of BP sites is recommended.
    • Older adults have increased systolic pressure related to decreased vessel elasticity.
    • Older adults often experience decreased BP after eating.
    • Instruct older adults to change position slowly and wait after each change to avoid postural hypotension and prevent injuries.

    HOME CARE CONSIDERATIONS

    • Assess the home noise level to determine the room that provides the quietest environment for BP measurement.
    • Teach the patient the importance of using an appropriate-size BP cuff for home BP measurement.
    • Assess the family’s ability to afford a sphygmomanometer for performing BP evaluations on a regular basis. Recommend electronic devices or aneroid sphygmomanometers that have proven to be accurate according to standard testing as well as appropriate-size cuffs.

    REFERENCES

    1. American Association of Critical-Care Nurses (AACN). (2016, reviewed 2021). AACN practice alert: Obtaining accurate noninvasive blood pressure measurements in adults. Critical Care Nurse, 36(3), e12-e16. doi:10.4037/ccn2016590
    2. American Heart Association (AHA). (2024). How to Manage High Blood Pressure. Retrieved July 2, 2024, from https://www.heart.org/en/health-topics/high-blood-pressure/changes-you-can-make-to-manage-high-blood-pressure
    3. American Heart Association (AHA). (2024). Home Blood Pressure Monitoring. Retrieved July 2, 2024, from https://www.heart.org/en/health-topics/high-blood-pressure/understanding-blood-pressure-readings/monitoring-your-blood-pressure-at-home#.Wuct7ExFwy9
    4. American Heart Association (AHA). (2024). Understanding blood pressure readings. Retrieved July 2, 2024, from https://www.heart.org/en/health-topics/high-blood-pressure/understanding-blood-pressure-readings#.Wucr9UxFwy9
    5. McLaughlin, S.A. and others. (2017). Considerations for clinicians in the diagnosis, prevention, and treatment of breast cancer-related lymphedema, recommendations from an expert panel: Part 2: Preventive and therapeutic options. Annals of Surgical Oncology, 24(10), 2827-2835. doi:10.1245/s10434-017-5964-6
    6. Muntner, P. and others. (2019). Measurement of blood pressure in humans: A scientific statement from the American Heart Association. Hypertension, 73(5), e35-e66. doi:10.1161/HYP.0000000000000087
    7. Pickering, T.G. and others. (2005). Recommendations for blood pressure measurement in humans and experimental animals. Part 1: Blood pressure measurement in humans: A statement for professionals from the Subcommittee of Professional and Public Education of the American Heart Association Council on High Blood Pressure Research. Hypertension, 45(1), 142-161. doi:10.1161/01.HYP.0000150859.47929.8e Retrieved July 2, 2024, from https://www.ahajournals.org/doi/full/10.1161/01.HYP.0000150859.47929.8e
    8. Schroeder, M.L. and others. (2024). Chapter 27: The child with cardiovascular dysfunction. In M.J. Hockenberry, E.A. Duffy, K.D. Gibbs (Eds.), Wong’s nursing care of infants and children (12th ed., pp. 949-1012). St. Louis: Elsevier.
    9. Sheppard, J.P. and others. (2019). Defining the relationship between arm and leg blood pressure readings: A systematic review and meta-analysis. Journal of Hypertension, 37(4), 660-670. Retrieved July 2, 2024, from https://journals.lww.com/jhypertension/Abstract/2019/04000/Defining_the_relationship_between_arm_and_leg.2.aspx

    ADDITIONAL READINGS

    Li, Y. and others. (2020). Effect of cuff positioning on the accuracy of blood pressure measurement with automated electronic blood pressure monitors. The Journal of Clinical Hypertension, 22(7), 1163-1172. doi:10.1111/jch.13902

    Adapted from Perry, A.G. and others (Eds.). (2022). Clinical nursing skills and techniques (10th ed.). St. Louis: Elsevier.

    Clinical Review: Martha Beck, MA, BSN, RN, CNOR

    Published: August 2024

    Small Elsevier Logo

    Cookies são usados neste site. Para recusar ou saber mais, visite nosso página de cookies.


    Copyright © 2024 Elsevier, its licensors, and contributors. All rights are reserved, including those for text and data mining, AI training, and similar technologies.

    Small Elsevier Logo
    RELX Group