ThisisClinicalSkillscontent

    Endotracheal Tube and Tracheostomy Cuff Care

    Learn more about Clinical Skills today! Standardize education and management competency among nurses, therapists and other health professionals to ensure knowledge and skills are current and reflect best practices and the latest clinical guidelines.

    Jan.30.2025

    Endotracheal Tube and Tracheostomy Tube Cuff Care - CE/NCPD

    The content in Clinical Skills is evidence based and intended to be a guide to clinical practice. Always follow your organization’s practice.

    ALERT

    If the patient is sedated, paralyzed, on pressure support, or receiving high oxygen delivery, do not disconnect the patient from the ventilator.

    Keep the cuff pressure at a level that maintains a seal between the cuff and the tracheal wall; the volume necessary to create the seal depends on tube size and cuff configuration.

    Potential complications of cuff inflation include tracheal stenosis, necrosis, tracheoesophageal fistulas, and tracheomalacia; these complications are more likely to occur in conditions that adversely affect tissue response to mucosal injury, such as hypotension.

    OVERVIEW

    The tube cuff helps stabilize the endotracheal or tracheostomy tube and maintains an adequate airway seal so that air moves through the tube into the lungs. The cuff is an inflatable “balloon” that surrounds the shaft of the tube near its distal end. When inflated, the cuff presses against the tracheal wall to prevent air leakage and pressure loss from the lungs. The goals of correctly inflating the cuff on an artificial airway are to promote lung inflation for mechanical ventilation and prevent aspiration of gastric contents, while at the same time allowing drainage of secretions that accumulate between the epiglottis and the cuff.

    The amount of air inserted in a cuff is based on several factors, including the size of the patient’s trachea and the external diameter of the artificial airway. If the cuff pressure is too high, permanent damage to the tracheal mucosa occurs.

    Appropriate cuff care helps prevent major pulmonary aspirations, helps prepare for tracheal extubation, decreases the risk of inadvertent extubation, provides a patent airway for ventilation and removal of secretions, and decreases the risk of health care–associated infections. Although a variety of endotracheal and tracheostomy tubes exist, the most desirable tube provides a maximum airway seal with minimal tracheal wall pressure, using a high-volume, low-pressure cuff (Figure 1)Figure 1.

    High-volume, low-pressure cuffs allow a large surface area to come into contact with the tracheal wall, thus distributing the pressure over a much greater area. The amount of pressure and volume necessary to obtain a seal and prevent mucosal damage depends on tube size and design, cuff configuration, mode of ventilation, and the patient’s arterial blood pressure.

    Cuff pressure should be measured once per shift or per the organization’s practice. Several devices are available to measure cuff pressure. A commercial electronic tracheal cuff pressure manometer can be attached directly to the inflation port to measure the cuff pressure (Figure 2)Figure 2. Other devices include bedside sphygmomanometers or special aneroid cuff manometers. Ideally, the cuff pressure should be between 20 and 30 cm H2O.undefined#ref1">1 Higher cuff pressure may compress tracheal capillaries, limit blood flow, and predispose the patient to tracheal necrosis. Lower cuff pressure may predispose the patient to aspiration of oropharyngeal secretions and the development of ventilator-associated pneumonia.1

    SUPPLIES

    See Supplies tab at the top of the page.

    EDUCATION

    • Give developmentally and culturally appropriate education based on the desire for knowledge, readiness to learn, preferred learning style, and overall neurologic and psychosocial state.
    • Provide the patient and family with an explanation of the procedure, the equipment, and the reason for tracheal tube cuff care.
    • Explain the patient’s role in assisting with cuff care.
    • Explain to the patient and family that the procedure may be uncomfortable and cause the patient to cough.
    • Encourage questions and answer them as they arise.

    ASSESSMENT AND PREPARATION

    Assessment

    1. Determine if the patient has health literacy needs or requires tools or assistance to effectively communicate. Be sure these needs can be met without compromising safety.
    2. Review the patient’s previous experience and knowledge of having an endotracheal or tracheostomy tube and understanding of the care to be provided.
    3. If the patient has an endotracheal tube, observe the position of the tube at the teeth, gum line, or naris.
    4. Assess for the presence of bilateral breath sounds.
    5. Assess for signs and symptoms of cuff leakage.
      1. Audible or auscultated inspiratory leak over the larynx
      2. The patient’s ability to vocalize audibly
      3. Deflation of inflation (pilot) valve balloon
      4. Loss of inspiratory and expiratory volume in a mechanically ventilated patient
    6. Assess for signs and symptoms of inadequate ventilation.
      1. Rising arterial carbon dioxide tension
      2. Chest–abdominal asynchrony
      3. Patient–ventilator asynchrony
      4. Dyspnea
      5. Headache
      6. Restlessness
      7. Confusion
      8. Lethargy
      9. Increasing (early sign) or decreasing (late sign) arterial blood pressure
      10. Activation of expiratory or inspiratory volume alarms on mechanical ventilator
    7. Assess the amount of air or pressure previously used to inflate the cuff.
    8. Assess the size of the tracheal tube and the size of the patient.

    Preparation

    1. Place the patient in the semi-Fowler position.

    PROCEDURE

    Measuring Cuff Pressure

    1. Suction the patient’s airway and oropharynx to clear secretions in the lower airway and above the cuff.
      Rationale: Clearing secretions decreases the risk of aspiration.
    2. Connect the commercial tracheal cuff pressure manometer to the inflation port of the endotracheal tube or tracheostomy tube according to the manufacturer’s instructions.
      Rationale: A manometer allows measurement of air pressure in the cuff.
      Adjust cuff pressure after attaching the manometer due to the evacuation of air volume from the cuff that may occur during attachment.
    3. Read the cuff pressure displayed on the manometer. Ideally, the cuff pressure should be between 20 and 30 cm H2O.1
      Keep the cuff pressure at a level that maintains a seal between the cuff and the tracheal wall; the volume necessary to create the seal depends on the tube size and cuff configuration.
    4. Disconnect the manometer line from the inflation valve.

    Correcting an Air Leak

    1. Suction the patient’s airway and oropharynx to clear secretions in the lower airway and above the cuff.
      Rationale: Clearing secretions decreases the risk of aspiration.
      1. Use a fresh, sterile catheter if using an open-suction system for suctioning the tracheobronchial tree. When suctioning of the tracheobronchial tree is complete, use the same catheter to suction the pharynx.
      2. If using a closed-suction system to suction the tracheobronchial tree, use a fresh, sterile catheter to suction the pharynx.
    2. Remove the T-tube (T-piece) or tracheostomy collar from the endotracheal or tracheostomy tube; or, if the patient is on a ventilator, is stable, and is not dependent on pressure support or high levels of oxygen, remove the ventilator tubing. Attach a self-inflating manual resuscitation bag (MRB) device for ventilation.
      If the patient is sedated, paralyzed, on pressure support, or receiving high oxygen delivery, do not disconnect the patient from the ventilator.
    3. Instill air into the cuff.
      1. Insert an air-filled syringe tip into the inflation valve (also referred to as the pilot balloon valve).
      2. Gradually add air to the cuff until the leak is eliminated.
      3. Ventilate the lungs with an MRB.
        Rationale: Ventilating the lungs reoxygenates the patient and allows for further assessment of proper cuff inflation and elimination of the air leak.
      4. If the patient is awake and alert, ask the patient to vocalize.
        Rationale: If the trachea is sealed and the air leak has been corrected, the patient will not be able to vocalize.
    4. Remove the syringe from the inflation valve.
    5. Reconnect the T-tube, tracheostomy collar, or ventilator tubing to the endotracheal or tracheostomy tube. Check and secure ventilator connections as needed.
      Rationale: Replacing oxygen and ensuring secure connections provides oxygen flow and prevents oxygen desaturation.
    6. Reassess the patient’s airway and respiratory status.

    Troubleshooting Cuff Problems

    Faulty Inflation Valve

    1. Check whether the cuff repeatedly deflates despite the addition of air.
      Rationale: A cuff that repeatedly deflates after air is instilled indicates a faulty inflation valve; repair is needed. There are commercial cuff inflation-valve repair kits available.
      If the inflation valve has become faulty and reintubation is undesirable, institute emergency cuff inflation (Figure 3)Figure 3.
      1. Insert a three-way stopcock into the distal opening of the inflation valve.
        Rationale: A three-way stopcock provides control of airflow in and out of the inflation valve.
      2. Inflate the cuff until an air leak is no longer detected.
        Rationale: If an air leak is no longer detected, this indicates that the seal between the tracheal wall and the cuff is restored.
      3. Clamp the inflation tube by applying a padded hemostat distal to the inflation valve.
        Rationale: Clamping the inflation tube maintains air in the cuff and provides a quick occlusion of the inflating tube.
      4. Turn the stopcock off to the inflation valve and leave it in place; remove the clamp.
        Rationale: Turning the stopcock off to the inflation valve allows temporary use of the tube while maintaining cuff pressure.
        Make plans to change the tube as a more permanent solution.
    2. Reassess the patient’s airway and respiratory status.

    Faulty Inflation Tube

    1. Check for an air leak in the inflation tube.
    2. If an air leak is found, cut off the faulty end of the inflation tube and valve with scissors.
      Rationale: Cutting off the faulty end prepares the inflation tube for repair.
    3. Insert a short, small-gauge blunt cannula into the inflation tube.
      Rationale: The cannula provides access for inflation.
      Use caution to avoid puncturing or severing the inflation line or skin.
    4. Attach a three-way stopcock to a blunt cannula.
      Rationale: A three-way stopcock provides control of airflow in and out of the inflating line.
    5. Inflate the cuff until an air leak is no longer detected.
      Rationale: Inability to detect an air leak after inflation of the cuff indicates that the seal between the tracheal wall and the cuff has been restored.
    6. Turn the stopcock off to the inflation tube.
      Rationale: Turning the stopcock off allows temporary use of the tube while maintaining cuff pressure.
    7. Secure the assembled device with tape to a tongue depressor.
      Rationale: Securing the device with tape to a tongue depressor provides stabilization and protection.
    8. Notify the practitioner and assemble the equipment for tube replacement.
    9. Reassess the patient’s airway and respiratory status.

    MONITORING AND CARE

    1. Assess respiratory status for optimal ventilation.
      Rationale: Inadequate interface between the tube cuff and tracheobronchial mucosa decreases inspiratory flow.
      Reportable conditions: Rising arterial carbon dioxide (CO2) tension, chest–abdominal dyssynchrony, patient–ventilator dyssynchrony, dyspnea, headache, restlessness, confusion, lethargy, rising (early sign) or falling (late sign) arterial blood pressure, activation of expiratory or inspiratory volume alarms on mechanical ventilator
    2. Measure cuff pressure every shift or per the organization’s practice. Maintain a cuff pressure between 20 and 30 cm H2O.1
      Rationale: Maintaining correct cuff pressure helps prevent tracheal injury and aspiration.
      If the volume of air (milliliters) needed to seal the airway increases, consult with the practitioner regarding the need for a chest radiograph to evaluate the patient for tracheal dilation and to determine the cuff diameter-to-tracheal diameter ratio. An increasing volume of air required to maintain cuff inflation may also indicate a leak in the cuff, inflation valve, or inflation tube.
    3. Minimize unnecessary tube manipulation to maintain cuff integrity.
      Rationale: Tube manipulation increases the likelihood of cuff disruption. A cuff leak or rupture is evident when the pressure on the manometer continues to decrease.
      Reportable conditions: Inability to maintain cuff inflation, audible air through the patient’s nose or mouth, low-pressure or low-volume alarm sounds on the mechanical ventilator, audible or auscultated inspiratory leak over the larynx, patient’s ability to vocalize audibly, pilot balloon deflation, loss of inspiratory and expiratory volume in a mechanically ventilated patient
    4. Suction secretions from the patient as indicated.
      Rationale: Removing secretions reduces the chance for partial or complete airway obstruction.
    5. Compare the patient’s cardiopulmonary status before and after cuff care.
      Rationale: Comparing the cardiopulmonary status before and after cuff care helps assess the effects of cuff care on the cardiovascular system.
      Reportable conditions: Decreased arterial oxygen saturation, cardiac arrhythmias, bronchospasm, respiratory distress, cyanosis, increased blood pressure or intracranial pressure, anxiety, agitation, changes in level of consciousness
    6. Reassess cuff pressure and volume during and after transporting the patient from one altitude to another (e.g., via air transport) or during hyperbaric therapy without environmental pressurization.
      Rationale: Changes in altitude alter the volume of gas in the cuff; therefore, volume and pressure must be reevaluated during and after transport.

    EXPECTED OUTCOMES

    • Endotracheal or tracheostomy tube remains in correct position.
    • Cuff pressure is kept at a level that maintains a seal between the cuff and tracheal wall.
    • Cuff remains intact.

    UNEXPECTED OUTCOMES

    • Extubation or tube dislodgment
    • Tracheal mucosal ischemia due to cuff overinflation
    • Faulty inflation valve or tube
    • Cuff overinflation and distention over the end of the tube
    • Cuff rupture

    DOCUMENTATION

    • Education
    • Date and time procedure was performed
    • Cuff pressure measurement
    • Cardiopulmonary and vital sign assessments before and after procedure
    • Patient’s tolerance of procedure
    • Appearance and characteristics of tracheal secretions, if present
    • Unexpected outcomes and related interventions

    REFERENCES

    1. Scott, J.B. (2025). Chapter 37: Airway management. In J.K. Stoller and others (Eds.), Egan’s fundamentals of respiratory care (13th ed., pp. 742-779). St. Louis: Elsevier.

    ADDITIONAL READINGS

    Satya Prakash, M.V.S., Aravind, C., Mohan, V.K. (2022). Comparative evaluation of three methods of endotracheal tube cuff inflation for adequacy of seal. Journal of Anaesthesiology Clinical Pharmacology, 38(4), 588-593. doi:10.4103/joacp.JOACP_560_20

    Adapted from Johnson, K.L. (Ed.). (2024). AACN procedure manual for progressive and critical care (8th ed.). St. Louis: Elsevier.

    Clinical Review: Genevieve L. Hackney, MSN, RN

    Published: January 2025

    Small Elsevier Logo

    Cookies are used by this site. To decline or learn more, visit our cookie notice.


    Copyright © 2024 Elsevier, its licensors, and contributors. All rights are reserved, including those for text and data mining, AI training, and similar technologies.

    Small Elsevier Logo
    RELX Group